6. Operating | 6.2 - Configuration and adjustment via the keypad | 91
5. Press / to make your selection.
6. Press / to select the digit positions.
7. Press / to increase/decrease the number.
8. Press "OK to confirm the input.
9. Press "Exit" to exit a menu.
If a parameter has been changed, SAVE appears on the display.
10. Press "OK".
yES appears on the display.
11. If necessary, press to select no.
12. Press "OK" to save the changes.
6.2 Configuration and adjustment via the keypad
6.2.1 Parameter table (setup)
 Cd   ·  Calibration data 
 000CALIb
nEW New calibration
VIEW View data
Mod Edit calibration
 001MAX.FSd
Position of decimal point, Max (maximum load/scale interval) and weight
unit
 002StEP
Scale interval
 003dEAdLo
byLoAd Set dead load with weight
byMV-V Set dead load with mV/V
 004SPAn
byLoAd Calibrate range with weight
byMV-V Calibrate range with mV/V
 005uVoLt.d
DisplayµV/d
 006WEIGHT
Display current gross weight, use the key for 10x resolution.