If problems occur when using the device, the following points should be checked.
SMART ONE can’t
connect with the
smartphone
The Bluetooth
connection is not
working properly
Look for SMART ONE on the list of recognized
devices. For correct use, the smartphone needs
Bluetooth version 4.0 or higher
The test results are
unreliable
Clean the turbine as described in the Maintenance
section. If necessary, replace the turbine with a
new one. If necessary, contact the manufacturer
The test was
performed wrongly
Repeat the test, following the directions on the
smartphone screen. Avoid sudden movements
when you finish exhaling. Discuss the value with
your physician
The turbine has not
been inserted properly
Insert the turbine from the front of the device by
pushing it all the way down and turning it
clockwise
11. ACCURACY AND RELIABILITY
This device meets the requirements of the following standard:
ATS Standardization of Spirometry 2005
Volume max 10 L
Volume accuracy 3.5% or 0.1 L whichever is greater
Peak Flow max 960 L/min (16 L/s )
Peak Flow accuracy 12% or 25 L/min (0.42 L/s ) whichever is greater