Smart One
ENGLISH
Rev.3.3 User Manual Page 25 of 299
15. INFORMATION FOR CORRECT USE IN AN ELECTROMAGNETIC ENVIRONMENT
Guidance and manufacturer’s declaration – electromagnetic emissions
The Smart One is intended for use in the electromagnetic environment specified below.
The customer or the user of the Smart One should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The Smart One uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Smart One is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The Smart One is intended for use in the electromagnetic environment specified below. The customer
or the user of the Smart One should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
In the event of disruption due to ESD during
oximetry test, the device recovers from any
disruption within 30 s. (according to ISO
9919).
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for
input/output
lines
Not applicable
Surge
±1 kV differential
Not Applicable