3. OPERATING PROCEDURE
3.2.3 Switching the analog range (analog resolution) during display of
“range”
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “range” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “range” from blinking to lighting-up
and blinks the currently-set analog range.
3 Press the [MODE] key quickly to switch the setting to a desired analog range.
Digital
resolution
Selectable analog range
0.0005mm ±0.01 ±0.02 ±0.05 ±0.1 ±0.2 ±40 ±80
0.001mm ±0.02 ±0.04 ±0.1 ±0.2 ±0.4 ±40 ±80
.00002” ±.0004 ±.0008 ±.002 ±.004 ±.008 ±2 ±4
.00005” ±.001 ±.002 ±.004 ±.01 ±.02 ±2 ±4
.0001” ±.002 ±.004 ±.01 ±.02 ±.04 ±2 ±4
RS-232
command *
A01 A02 A05 A10 A20 AFS AFL
*: For information about RS232 commands, refer to section 4.2.4.
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “dir”.
3.2.4
•
Switching the count direction during display of “dir”
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “dir” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “dir” from blinking to lighting-up. Only
“dir” is displayed if the currently-set direction is “+” count, and “REV” blinks at the left
of “dir” if the direction is “-" count.
3 Press the [MODE] key quickly to switch the setting to a desired direction (between +
and -).
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “output”.
NOTE
Upon change of the count direction, the sign (+/-) of a display value is also changed.
Before calling a preset value, perform setup of the count direction.
3.2.5
3.2.5.1
Switching the I/O format between SPC(digimatic) and RS232 during
display of “output”
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “output” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “output” from blinking to lighting-up
and blinks the currently-set I/O format.
3 Press the [MODE] key quickly to switch the setting to a desired I/O format (between
SPC and RS232).
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “id” (section 3.2.6) if SPC is specified. If RA232 is specified as the I/O format,
the procedure is continuously switched to the RS232 setup (section 3.2.5.1).
RS-232 setup
1 When the procedure is switched to the RS232 setup, “bps” starts blinking.
2 Press the [SET/ZERO] key quickly to blink the currently-set baud rate.
3 Press the [MODE] key quickly to switch the setting to a desired baud rate (4800 or
9600).
4 Quickly press the [SET/ZERO] key to apply the communication setting and switch the
setup item to the next “p-bit.”.
5 Referring to the following table, set the parity bit and data bit according to the
No. 99MAH016B
3-3