3.3.2 Switching a measurement type in the measurement mode
This instrument is provided with measurement mode described in the table below in
addition to normal measurement mode.
Press the [MODE] key quickly to switch the measurement mode to a desired
measurement type.
Reference Measurement mode Display Setup content
--- Normal Performs normal measurement.
3.3.2.1 Tolerance judgment
“<○>”
Performs tolerance judgment
measurement.
Tolerance judgment is
performed for each hold value
in the runout, maximum value,
and minimum value hold
measurements.
3.3.2.2 Runout peak hold
“TIR”
Performs peak hold
measurement of the runout
(total indicator reading).
3.3.2.3 Maximum value peak
hold
“Max”
Performs peak hold
measurement of the maximum
value.
3.3.2.4 Minimum value peak
hold
“Min”
Performs peak hold
measurement of the minimum
value.
3.3.2.1 Tolerance judgment (setting/checking the upper and lower limit values)
Tolerance judgment is performed for the current value in the normal measurement
mode, whereas it is performed for each peak hold value in each peak hold mode.
It is possible to set the upper and lower limit values separately for individual INC
system and Preset system.
It is not possible to set the upper and lower limit values separately for each
measurement mode.
The tolerance judgment is displayed with “< c >” and the backlight (red for NG and
green for OK).
1 Quickly press the [MODE] key repeatedly until “< c >” on the LCD starts blinking.
2 When pressing the [SET/ZERO] key quickly, symbol “off” or “on” starts blinking.
3 Press the [MODE] key quickly to switch the blinking symbol to “on”.
4 Press the [SET/ZERO] key quickly. Then, the most recently-set upper limit value is
displayed and the “>” symbol starts blinking.
5 Quickly press the [SET/ZERO] key again to apply the upper limit value and switch the
setup item to the lower limit value.
To change the upper limit value, press and hold the [SET/ZERO] key for a long time to
blink the “+” or “-” symbol, entering the state where the upper limit value can be set.
No. 99MAH016B
3-6