3.3.2.4
•
Minimum value peak hold during display of “Min”
Hold the minimum value of fluctuating measurement values according to the following
procedure.
When the tolerance judgment condition has been set, the instrument displays the
tolerance judgment result for the measured minimum value.
1 Quickly press the [MODE] key repeatedly until “Min” is displayed.
2 Press the [SET/ZERO] key shortly to blink the “Min” symbol. Then, the minimum value
being held is canceled and a new minimum value hold measurement starts.
3 When the spindle is displaced, the “Min” symbol changes from blinking to lighting-up
and the minimum value is held.
NOTE
Press and hold the [SET/ZERO] key for a long time during measurement in the Preset
system. Then, the “INC” symbol lights up and the preset position is canceled. It is
possible, thereby, to perform measurement on the basis of the minimum value.(peak
zero set)
TIP
• The difference of display and tolerance judgment in each measurement mode is shown
below.
Difference of display in TIR, Max., Min. mode
0
→
a
→
b
→
c
→
d
Normal
0.00 5.00 -5.00 10.00 0.00
TIR 0.00 5.00 10.00 15.00
Max. 0.00 5.00 10.00
Min. 0.00 -5.00
Difference of tolerance judgment in each mode(Upper limit 8.00, Lower limit -3.00)
0 a b c d
Normal
TIR
Max.
Min.
c
10.00
a
5.00
d
0.00
0
Lower limit
-3.000
-5.00
b
Upper limit 8.000
No. 99MAH016B
3-8