5 ERROR MESSAGE AND
ACTION
5
This chapter describes the error messages and their
corrective actions.
If any error occurs in this Digimatic indicator, a corresponding error message is
displayed. (While RS-232 communication is being performed, an error message is
also outputted. In the Digimatic code output mode an error message will not be
outputted.)
If the instrument does not recover the normal conditions after corrective actions,
contact your dealer or Mitutoyo sales office.
1. Errors possibly caused in normal operation
Message
RS-232
output
Definition Action
Error20 Over-speed
• •
•
The spindle operation
speed has exceeded
the response speed of
this instrument.
Press the [SET/ZERO] key.
Use this instrument at a speed so that
the spindle operation speed is within the
response speed.
Error30 Display overflow
• •
•
•
The display value has
exceeded the
maximum number of
digits that can be
displayed.
Return the display value to the
permissible number of digits to
automatically reset the error.
Press the [SET/ZERO] key quickly or
hold it down slowly. *1
Enter the setup mode with the [MODE]
key, and then reduce the resolution. *2
(Normal
count)
Error52 RS-232 command input
error
•
•
A wrong command
was inputted.
Re-enter the correct command.
Error90
Tolerance setup error
•
•
The limit values are set
so as to be an upper
limit value < a lower
limit value.
Press the [SET/ZERO] key quickly, and
then set tolerance limit values so as to
be an upper limit value > a lower limit
value.
Error95P Preset value setting
error
•
•
•
The preset value has
over flown.
Set a preset value again. *3
Reduce the resolution. *3
Error95G Upper limit value setting
error
•
•
•
The upper limit value
has over flown.
Set an upper limit value again. *4
Reduce the resolution. *3
Error95D Lower limit value setting
error
•
•
•
The lower limit value
has over flown.
Set a lower limit value again. *4
Reduce the resolution. *3
*1. Press the [SET/ZERO] key quickly (less than 2 seconds) to switch to the
Preset Setup mode, or press and hold the key slowly (2 seconds or
more) to perform zero-setting in the INC measurement system.
No. 99MAH016B