No. 99MAH016B
4-8
3 Example of communication
The procedure below is the case when the tolerance judgment is performed after presetting
+12.3455mm and setting the upper limit for 12.35mm and the lower limit 12.34mm in the
preset system. For this time, ID No. is “00” (default).
(1) Perform
presetting: Input the command “CP00,+0012.3455” from PC. This
instrument performs presetting and responds the command “CH00”.
(2)
Set Upper limit: Input the command “CG00,+0012.3500” from PC. This instrument
sets the upper limit and responds the command “CH00”.
(3) Set
Lower limit: Input the command “CD00,+0012.3400” from PC. This instrument
sets the lower limit and responds the command “CH00”.
(4)
Set the tolerance judgment mode. Input the command “DJ00,ON” from PC. This
instrument changes the measurement mode to the tolerance judgment mode and
responds the command “DH00”.
TIP
For the example of RS-232 communication described above, perform presetting (change of
the measurement system) at the beginning. This instrument changes the upper and the
lower limit of the measurement mode when inputting the command.
For the procedure (2) to (4), it is allowed to proceed with the random order.
When using the hyper terminal, it is not possible to correct the command typing error. Once
input the incorrect command (this instrument responds the command “
CH**,Error 52”)and
then input the correct command.