No. 99MAH016B
5-2
*2. Press and hold the [MODE] key slowly to switch to the Parameter Setup
mode.
*3. Press the [MODE] key quickly to clear the error using the procedure in *1
and *2 “ Error 30 message”.
*4. Press and hold the [MODE] key slowly to blink the “<c>” symbol and
make tolerance setup ready, or press and hold the [SET/ZERO] key
slowly to display a maximum permissible display value and set tolerance
limit values.
NOTE
If data input/output is not performed properly, there is a possibility of an error
in communication setup. Check the settings of this instrument and the
connected device.
For detailed information about the RS-232 output command, refer to section
4.2.
If any of the above errors occurs during function lock, press and hold the
[DATA/Fn] key for 2 seconds or more to cancel the function lock, and then perform
the error clear processing.
2. Errors possibly caused during the use of Presetter
Message Definition
Action
Communication error
A value sent from the
Presetter has exceeded
the permissible range of
the instrument.
Press the [SET/ZERO] key quickly.
Set the value again so that its decimal point
position is consistent with that of the display
value on this instrument.
Tolerance setup error
The limit values are set so
as to be an upper limit
value < a lower limit value.
In a wait state of lower
limit value entry, a preset
value or upper limit value
was inputted.
Before setting an upper
limit value, a lower limit
value was set.
Press the [SET/ZERO] key.
Re-set limit values so as to be an upper limit
value > a lower limit value.
Perform re-setting in the correct procedure.
NOTE
For detailed information about the Presetter, refer to the Presetter User’s Manual.