command.
A Digimatic indicator with the ID No. of 00 can receive and execute a command irrespective
of specification of any ID No.
After power is turned on and moves to the preset system for the first time, the
communication command of “DH**, PRESERCRLF” is indicated (at this time, the preset
value is not determined). Continue to input the preset value.
When performing the setting of the preset value in the INC system, the measurement mode
is moved to the preset system automatically.
When moving from the INC system to the preset system, each peak hold mode is canceled.
Specify a preset value and an upper/lower limit value with 10 digits of sequence including a
sign and a decimal point. However, the number next to the sign is fixed to “0”. A display
value is also outputted with the same sequence.
For the presetting and the tolerance value setting, input the value and the decimal point
position with respect to the resolution.
"Error 90" judgment is performed by comparing the new tolerance value with the current
tolerance value.
After receiving the output in response to a command, transmit a next command. If there is
no response to a command, clear the communication buffer, then transmit the command
again after a lapse of one second or more.
This instrument executes key operation on a top-priority basis. The instrument temporarily
stops the RS-232 communication function during key operation, and then executes
command and data output upon restoration of the countable state.
2 Parameter setup command
Operation content
Input
(External device -> this
instrument)
Output
(This instrument -> external
device)
mm DU**,MMCRLF
inch
DU**,INCRLF
DH**CRLF
Status output
(mm)
DH**,MMCRLF
Switching unit
Status output
(inch)
DU**,OUTCRLF
DH**,INCRLF
0.0005mm
DR**,D0.000500CRLF
0.001mm
DR**,D0.001000CRLF
.00002”
DR**,D0.000020CRLF
.00005”
DR**,D0.000050CRLF
.0001”
DR**,D0.000100CRLF
DH**CRLF
Switching
resolution
Status output
DR**,DOUTCRLF DH**,D0.000500CRLF
Range setting
DR**,A##CRLF DH**CRLF
Switching
analog range
Status output
DR**,AOUTCRLF DH**,A##CRLF
+ direction
DD**,NORMCRLF
- (rev.) direction
DD**,REVCRLF
DH**CRLF
Status output (+)
DH**,NORMCRLF
Switching
count
direction
Status output (-)
DD**,OUTCRLF
DH**,REVCRLF
Start setting (INC)
DS**,SINCRLF
Setup when
startup
Start setting (Setup
during stop)
DS**,SFREECRLF
DH**CRLF
INC / INC start
DH**,INC‑SINCRLF
INC / start during
stop
DH**,INC‑SFREECRLF
P / INC start
DH**,P‑SINCRLF
Measurement
system/Setup
output when
startup
P / start during
stop
DS**,OUTCRLF
DH**,P‑SFREECRLF
Return initial setting
DE**,RESETCRLF DE**,RESETCRLF
No. 99MAH016B
4-6