EasyManua.ls Logo

MKS 670B - Diagnostic Messages; Table 24: RS-232 Diagnostic Messages

MKS 670B
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RS-232 Messages Cha
p
ter Six: RS-232 Communications O
p
tion
98
Diagnostic Messages
The Diagnostic messages, listed in Table 24, consist of one command (80) to provide diagnostic
information. The <id> field identifies the diagnostic information requested.
Diagnostic Messages
Parameter Entry
Id Data
Response
80 0 = Request Answer Back ? *U*U*U (produces an
alternating bit pattern)
1 = Request Version Number ? Version number
2 = I/O Connector Response 8-bit output
image
?
Produces the 8-bit output
image at the I/O connector
16-bit input image
3 = Report Actual Gain Range ? 0 = x1
1 = x 0.1
2 = x 0.01
Table 24: RS-232 Diagnostic Messages
Example 1: How To Check the Version Number
To check the version number, send the following query:
@801?
ENTER
The 670 instrument will respond with:
@
801X.X
where
X.X
represents the actual version number.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents