IEEE-488 Device Compliance
Refer to Table 27 for a description of the IEEE-488 device compliant functions.
IEEE-488 Device Compliance
Function Command Compatibility
Source Handshake SH1 Complete
Acceptor Handshake AH1 Complete
Talker T6 Talker, serial poll and unaddress if my-listener-
address (MLA)
Listener L4 Listener and unaddress if my-talker-address (MTA)
Service Request SR1 Can request (asynchronously) service from controller
Remote Local RL0 No capability for lockout
Parallel Poll PP0 No capability to send message without being
addressed to talk
Device Clear DC1 Complete
Device Trigger DT0 No Capability
Controller C0 Does not have the capability to send commands or
addresses to other devices
Talker TE0 Talker function with address extension not available
Listener LE0 Listener function with address extension not
available
Electrical Interface E1 Open collector
All the functions listed in this table are bus functions
Table 27: IEEE-488 Device Compliance
Buffer Length
The buffer length is fixed at 256 characters.
Address
Every device on the IEEE-488 bus must have an address assigned to it. To set the address of 670
instrument, scroll to the IEEE-488 configuration screen, under the [
SETUP
] key menu. The
address entry ranges from 1 to 30. The 670 instrument is initially configured with the address set
to 8.