41
EN
TECHNICAL MANUAL - B CLASSIC / B FUTURA / S CLASSIC
ALARMS
S 032
LCD Indication Alarm description Effect / Symptom
SYSTEM ERROR
Check through watchdog that one of the main
tasks is not blocked (e.g. innite loop)
Possible causes / Checks Solution
Malfunction of the control
software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service
S 040
LCD Indication Alarm description Effect / Symptom
LOG ERROR Check log saving in the ash memory
Possible causes / Checks Solution
Malfunction of the control
software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service