ALARMS
S 011
LCD Indication Alarm description Effect / Symptom
PRINTER: DOOR OPEN
A failed/bad closing of printer
door was detected.
Possible causes / Checks Solution
Door open.
Roll improperly positioned.
Check closing.
Check positioning.
S 012
LCD Indication Alarm description Effect / Symptom
PRINTER NOT READY
A failed/bad set-up of printer was
detected.
Possible causes / Checks Solution
Improper set-up. Set up.
S 020
LCD Indication Alarm description Effect / Symptom
PERFORM BACK-UP Request cycle report download. Message is displayed at every cycle.
Possible causes / Checks Solution
250 cycles have been
run and report was never
downloaded.
Download cycle report, in PDF format, using “NEW” item from menu DATA
MANAGEMENT/USB.
S 021
LCD Indication Alarm description Effect / Symptom
DATA OVERWRITING
(Classic only)
Warning before starting to
overwrite cycle report data.
Message is displayed at every cycle.
Proceeding without downloading will delete any
overwritten cycle.
Possible causes / Checks Solution
500 cycles have been
run and report was never
downloaded.
Download cycle report, in PDF format, using “NEW” item from menu DATA
MANAGEMENT/USB.
S 030
LCD Indication Alarm description Effect / Symptom
SYSTEM ERROR
Check through watchdog that one of the main
tasks has not crashed
Possible causes / Checks Solution
Malfunction of the control
software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service
S 031
LCD Indication Alarm description Effect / Symptom
SYSTEM ERROR
Check through hardware watchdog that one
peripheral device is not blocked
Possible causes / Checks Solution
Malfunction of control board
or software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service