47
EN
TECHNICAL MANUAL - B CLASSIC / B FUTURA / S CLASSIC
SERVICE MENU
SQ11
From the DOOR menu, you can establish:
- door state (closed/locked)
- the states of closing system microswitches
- the motor current draw values during the different door closing phases
Door release, unlock and complete opening can be activated manually.
COMPONENT TEST – FUNCTIONAL CHECKS –
SOLENOID VALVES/OTHER COMPONENTS
SQ12
The following menus can be used:
- SOLENOID VALVES
- OTHER COMPONENTS
These menus are used to test individual component or to activate and test
several components at the same time to check the different functions that
are supposed to be performed during a sterilization cycle.
To activate several components available in the two menus at the same
time, you may toggle between the two menus using the command
, which changes between menus while retaining the previously selected
commands.