B CLASSIC / B FUTURA / S CLASSIC - TECHNICAL MANUAL
SERVICE MENU
SQ18 – CHECK FOR WATER INJECTION PUMP/EV6 ACTIVATION
Keeping the autoclave door open, from menu OTHER COMPONENTS:
- Activate WP – injection pump/EV6
- Check that water is supplied to chamber properly
Perform test only if PT2 - steam generator temperature is lower than 70°C.
De-activate the components after completing the test.
SQ19 – CHECK FOR TUBE BUNDLE HEATING ELEMENT ACTIVATION
From the OTHER COMPONENTS menu, activate:
- BH - chamber tube bundle heating element
Proper activation is indicated by parameter PT3 – chamber tube bundle
heating element temperature increasing.
SQ20 – CHECK FOR GENERATOR HEATING ELEMENT ACTIVATION
From the OTHER COMPONENTS menu, activate:
- SG – generator heating element
Proper activation is indicated by parameter PT2 – steam generator heating
element temperature increasing.