B CLASSIC / B FUTURA / S CLASSIC - TECHNICAL MANUAL
SERVICE MENU
MENU SERVICE – 134°C STERILIZATION
SQ36
Select 134°C STERILIZATION to display a screen where it is possible to
select the sterilisation time between 4 and 5.5 minutes.
This option is valid only for sterilisation cycles performed at 134°C, except
for Prion cycle.
Conrm this selection with ENTER.
MENU SERVICE – EMPTYING CYCLE
SQ37
The EMPTYING CYCLE can be started after manually emptying the tanks.
Vacuum pump and heating system are activated with the aim to remove
the water and humidity left in the device.
Select EMPTYING CYCLE to display a screen where it is possible to
activate the complete emptying of the autoclave and set the start delay.