Appendix G: Maintenance and Service Instructions
Inspection
Inspection
Inspection should proceed as follows.
1) From the Diagnostics menu, perform the Front Panel Check to verify the front panel controls.
Each of the front panel controls is represented on-screen. See the section on the Diagnostics
menu for more information.
2) Remove any connectors from the CT100.
3) From the Diagnostics menu, Main Board submenu, perform the Memory, GPIO, and
Peripherals check. See the section on the Diagnostics menu for more information.
4) From the Diagnostics menu, perform the Analog check. See the section on the Diagnostics
menu for more information.
5) Use the instructions in Appendix C of this manual to perform a calibrations check.
Inspection interval
Inspection interval
The CT100 should be inspected every 6 months.
Calibration and calibration interval
Calibration and calibration interval
See Appendix D for the appropriate calibration interval and detailed instructions on how to
calibrate a CT100.
108 Mohr CT100 / CT100HF Operator's Manual