Appendix D: Calibration Procedures
Clear temperature lookup table
Clear temperature lookup table
After calibration is completed, it is necessary to clear the temperature lookup table used to
expedite temperature calibration while the CT100 is in use. Clearing this table will force the
temperature lookup table to be rebuilt using the newly calibrated values.
To clear the temperature lookup table:
1) From the main screen, push the blue MENU button.
2) Enter the Calibration submenu (Button 4).
3) Select OK to continue (Button 4).
4) Select Manual Calibration (Button 21).
5) Select Clear Temp Lookup by pressing Button 23. A warning message will be displayed
asking to verify the clearing of temperature adjustment.
6) Select OK to complete clearing the temperature lookup table.
Mohr CT100 / CT100HF Operator's Manual 89