Operating Instructions
Hybrid adjust info
Show the correction rate and reference temperature for the Hybrid Comp
calibration temperature adjustment.
Temperature adjustment
When temperature adjustment is on, the CT100 will correct measurements for
temperature variations and the CT100 will from time to time pause to adjust to a new
temperature. Temperature adjustment should be turned off when calibrating, except
when setting horizontal calibration, in which case temperature adjustment should be
turned on.
Vertical calibration
Vertical calibration dialog for all three cable lengths.
Driver start auto calibrate
Step-by-step process with on-screen directions.
Capacitive load auto calibrate
Step-by-step process with on-screen directions.
Resistive load auto calibrate
Step-by-step process with on-screen directions.
Horizontal calibration variables
This sets the cursor's position as the zero point for distance measurements. This is
normally the first falling edge when nothing is connected to the front cable connector
(second rising edge on CT100HF units).
Info menu
This menu provides more information regarding the instrument and its installed software.
License info
This option gives information about the license installed on the CT100, including the
license date and enabled options.
License from USB
Use this option to load a license code from a USB flash drive. Detailed instructions will
be provided with the license code.
Web update
When a CT100 is connected to the Internet, this item will automatically update the
CT100 if newer software is available.
Enter license
As an alternative to loading a license code, it can be manually entered using this option.
This is useful for operators who are unable to connect to a computer or use a flash
drive.
Software version
Display software and firmware version information. This also serves to verify the
communications links between the micro-controllers.
30 Mohr CT100 / CT100HF Operator's Manual