Appendix C: Operator Performance Check
should appear on the device display. Switch the CT100 into the short pulse setting from the Main
menu. Measure the pulse frequency coming off the front panel cable connector. Repeat pulse
frequency measurements for short, medium, and long pulse settings. The pulse frequency should
be 250 kHz for short, 50 kHz for medium, and 12.5 kHz for long.
78 Mohr CT100 / CT100HF Operator's Manual