2-7
TimeRemaining(checkbeforeeachday’suse)
In normal operating mode, the Altair Detector’s remaining
operating time (0 to 24 months) is always displayed.
When less than one month remains, the display changes to the
number of days remaining.
The instrument is designed to continue to operate after 24 months
have elapsed. A "+" followed by a numerical value and "months"
or "days" indicates that more than 24 months have expired.
For the instrument to function properly (including after the
24-month initial period has elapsed), the end user must continue
to perform an alarm function check and bump test before each
day’s use.
The instrument will continue to operate as long as:
The Low Battery warning indicator is not displayed
The instrument continues to pass the bump test as outlined in
this manual.
For the oxygen version:
If, after the initial 24-month period, the sensor output is less than
5% O
2
for more than five minutes, then it is considered a sensor
failure:
Instrument will display "SNS/ERR".
If the battery reaches its end of life, the
LowBatterywarning indicator activates:
User must discontinue using the Altair Single Gas
Detector at this time, although the instrument is still
detecting gas.
Low Battery indicator flashes
Months remaining continues to display.
LowBatteryalarm activates:
Horn sounds
LEDs flash
Low Battery indicator flashes
"ERR" displays
Pressing the "TEST" button silences the alarm
Unit is no longer detecting gas and must be taken out of
service.