EasyManua.ls Logo

MSA Altair - Page 21

MSA Altair
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Press the TEST button again:
The "" that appears on the display:
indicates that the instrument passed the bump test.
remains for 24 hours, indicating that a bump test
was performed on this instrument within the last
24 hours.
NOTE: See FIGURE 3-1 for details
If If the "" does not appear and and "ERR" is indicated:
Check that the:
sensor inlet is not clogged
correct calibration cylinder is used to perform the
bump test
gas cylinder has not expired and is not empty
gas was applied at the appropriate time
gas tubing is seated in the front instrument case
during testing.
Repeat the bump test process as needed.
If the "" does not appear,calibrate the instrument and repeat the
Bump Test.
3-2

Table of Contents

Other manuals for MSA Altair

Related product manuals