EasyManua.ls Logo

MSA Altair - Page 29

MSA Altair
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Press the TEST button at the "FAS" "?" screen to perform a
calibration at 20.8% O
2
.
NOTE: This procedure must be performed in fresh,
uncontaminated air. Do not breathe on the sensor while
performing this function.
If the successfully calibrates:
"OK" displays
If the sensor does not successfully calibrate:
"ERR" displays.
4. Wait five seconds.
The instrument returns to normal operating mode.
5. If "ERR" displays after calibration, the current settings did not
change. Immediately check that:
The instrument is in fresh, uncontaminated air during the
calibration process
No one breathes on the sensor during calibration.
Repeat steps 1 through 6, as necessary.
6. The display must read "OK"; if "ERR" remains, remove the
instrument from service.
7. Perform a bump test to confirm operation and activate the "".
4-7

Table of Contents

Other manuals for MSA Altair

Related product manuals