15. “Probe Failure”
Use this menu to assign a maximum time for connected probes to stay in “stuck” condition. A stuck probe (it
remains on same reading value for some time) means that probably probe itself is damaged. Within this menu
is possible to STOP pumps or just to show an alarm message (probe failure) . Function can be disabled selecting
“OFF” instead of a number (minutes). This function can be set for both or just one probe.
E.g. To set CD pump to stop after time ends and probe doesn’t change reading values press wheel, choose maxi-
mum time, press wheel move on next field and choose “STOP”. Time can be set between 100 and 254 minutes.
When satisfied with settings move on exit and press wheel.
To end procedure move cursor on “OK” and press wheel to proceed to “Save” request screen. Move wheel on “YES”
to save or “NO” to discard changes.