EasyManua.ls Logo

Oakton CON 11 - P8.0: Selection of Cell Constant

Oakton CON 11
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual CON 11/110
50
8.9 P8.0: Selection of cell constant
Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements (below 20
µS or 10 ppm).
The cell included with your meter has a cell constant of K =
1.0.
From measurement mode
1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view
parameter P8.0.
3. Press the HOLD/ENTER key to enter parameter
P8.0.
4. Press the MI/▲ or MR/▼ keys to select the cell
constant between K = 1.0, 0.1, or 10.
5. Press the HOLD/ENTER key to confirm selection
and to return to the subgroup menu.
6. Press the CAL/MEAS key to return to measurement
mode.
cel
1.0
SETUP
SETUP
p 8.0
p 8.0
p 8.0
10.0
SETUP
MI
MR
HOLD
ENTER
K =
K =
Figure 43: P8.0-
Selecting Cell Constant

Table of Contents