CHAPTER 2 SEVICE MENU
Ohaus Corporation www.ohaus.com 2-8 FC5706 Centrifuge Service Manual
Figure 11
Press the button "rpm/rcf" (4). The display "rpm/rcf" (A-1) is now activated and you can
choose the respective setup between -30 and 20 with the potentiometer (8).
Note: After all points have been reviewed and/or set up, you have to press
shortly the key "start" to store all parameters. Otherwise the setups were all for
nothing. If the storage has been successful in the display „rpm/rcf“ the value of
the imbalance sensor and the word „StorE“ are indicated.
Figure 12