EasyManua.ls Logo

Oldham iTrans

Oldham iTrans
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iTrans Fixed Point Single/Dual Gas Monitor Operation
(P/N: 77023554-1) iTrans User Manual 5-11
The 4-20 mA setpoint is designated
with a 4 displayed in status bit and
the current high end range next to it.
To change the range, press the
button during the 10-second
countdown. If you do not press
during the 10-second countdown, the
will return to the Normal
Operating Mode. If you initiate the
4-20 mA range option, the status bit
will start to flash and the
range setpoint can be changed by
using the “” and “” keys.
When the desired value is reached,
press the key. If the value is not
saved before the time-out, the
will go back to the
Programming Mode.
Figure 5-18. Changing the Analog
Output Upper Value
5.6.6. Set System Time Minute
The system’s clock minute setting is
designated with a “1” in the status
bit and current value next to it. To
change the minutes, press the
button during the 10-second
countdown. If you do not press
during the 10-second countdown, the
will return to the Normal
Operating Mode. If you initiate the
minutes option, the status bit will
start to flash and the
minute can be changed by using the
” and “” keys.
Figure 5-19. Setting System Time
(Minutes)

Table of Contents

Related product manuals