2.5.3 Laser Power Selector, Figure 7.3.
These controls regulate the percentage of laser power allowed to pass through the acoustico-optical
tunable filter (AOTF) located in the scanhead. Each individual laser has a unique output power. Thus, if
all laser lines are set at any given percentage, each will be providing a different amount of energy to the
sample.
2.5.4 Scroll List, Figure 7.4.
If there are more lines than can be displayed in the window, scroll through the list by clicking on the
black triangles or by dragging the slider.
2.5.5 Line Settings, Figure 7.5
As with other controls, the laser power may be adjusted by clicking anywhere in the control area,
scrolling the thumbwheel (1% increments) or by clicking on the black triangles (0.1% increments).
Using the thumbwheel to go below 1% will set the laser power to 0%. Use the triangles to increment
laser intensities to values between 0% and 1%.
Figure 7. Selecting laser lines and setting laser power
7.1 Laser Line
Selection
7.3 Line Power Selector
7.2 Laser Line
7.5 Line Settings
7.4 Scroll List of Lines
Olympus Fluoview-1000 User’s Guide
V.M. Bloedel Hearing Research Center, Core for Communication Research
Center on Human Development and Disability, Digital Microscopy Center
May 11, 2011 15