EasyManua.ls Logo

Olympus FV3000 - Lambda Series Acquisition

Olympus FV3000
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lambda series
(acquiring by a single channels)
*Before starting the following procedure,
make adjustments for imaging.
(refer to page 4-8.)
Changing to Lambda mode
Select “Lambda” in [Mode] in [PMT
setting] Tool Window.
Select one channel to be used
Press the button. The
[Dye & Detector Select] dialog box
appears. Press the button to reset
the Assigned Detector.
Select the [Channel] tab in [Dye &
Detector Select] dialog box. Select the
“High Sensitive Detector CH1” in
[Detector List], and press button.
Press the button to close the [Dye &
Detector Select] dialog box.
Tick the checkbox of the wavelength of
the excitation laser to be used.
Setting DM and SDM
Select [Lightpath] in [Tool Window]
menu. [Lightpath] Tool Window appears.
Select [LSMScanner] tab in [Lightpath]
Tool Window. Press the DM button to
display the dichroic mirror list. Select
“DM” which reflects the wavelength of
the excitation laser selected in ④.
Check the light path whether the light
path is set properly. See the figure(A).
(SDM→MIRROR→MIRROR→HSD1)
Setting “Band width” and “Step Size”
In [PMT Setting] Tool Window, set the
width of the wavelength for which the
spectral image acquisition is performed
per each step in [Band width], and set
the interval between the wavelength to
start acquisition in the next step in [Step
size].
Standard
・Bandwidth
15μm is recommended. If the live image
is dark, widen the Bandwidth.
・Step Size:
5nm is recommended. The narrower the
Step Size, the more accurate the analysis.
Figure(A)
26

Other manuals for Olympus FV3000

Related product manuals