UM - EHY2000plus-R24
66
Take care that during this process, air must not get into the bolus. Take care
that you can continuously see the water in the filling hole. Now, in this position
screw in the packing nut.
7. Turn up the electrode and check the appropriate volume of the water
in the bolus and that it is de-aerated once more.