Attention! In the next pictures we are showing some examples of non-proper
filling up of the electrodes and we are also showing how to avoid the
mistakes.
Air must not remain in the electrode because this blocks or misleads the
tuning and the auto matching procedure and so suppresses the efficacy of
the treatment. To correct this mistake, go back to point 5.
.
By pushing the bolus with your fingertips you cannot reach the metallic plate
of the electrode: the electrode is overfilled. In this case the electrode cannot
fit well on the patient’s treated region, the tuning accuracy decreases.
Reduce the amount of water in the bolus! To correct this mistake, go back to
point 6.