Opticon
OPR 3201
Specifications Manual
38
14.3. Vibration Strength (without packaging)
No malfunction occurred after the following vibration test.
Vibration test: Increase the frequency of the vibration from 10 Hz to 100 Hz with
accelerated velocity 19.6m/s
2
(2G) for 30 minutes (60 minutes for one cycle) in non-
operating state. Repeat this routine in each X, Y, Z directions.
14.4. Vibration Strength (with individual packaging)
No malfunction occurred after the following vibration test.
Vibration test: Increase the frequency of the vibration from 10 Hz to 100 Hz with
accelerated velocity 19.6m/s
2
(2G) for 30 minutes (60 minutes for one cycle) in non-
operating state. Repeat this routine in each X, Y, Z directions.
14.5. Dust and Drip Proof
IEC IP42
Dust Prevention
Level Details
4 Prevention of objects larger than 1 mm. Most wires, screws, etc.
Water Prevention
Level Details
2
Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up
to 15° from its normal position.
14.6. Cable Strength
No malfunction occurred after the following cable strength test.
Affix the scanner to an immovable object, then pull it with a static load of 24.5 N (2.5 kgf)
for 1 second. Carry out this test 20 times.