7.4.3 Attendant control fault overview
Battery LED
LED Fault/Warning Cause Possible corrective action
Red is flashing Battery undervoltage Battery deep discharge Charge as soon as possible
Green is flashing Battery overvoltage Voltage too high (after com
plete charging and driving
downhill)
Continue driving slowly
Mode LED
LED Fault/Warning Cause Possible corrective action
Red is flashing Various product faults More detailed information about the existing fault is shown
on the LCD monitor on the control panel (Wheelchair con
trol unit fault overview f.).
Seat function LEDs
LED Fault/Warning Cause Possible corrective action
Backrest LED is flashing Back angle adjustment motor
fault
Faulty cabling or plug con
tact
Defective actuator
Check cabling/plug connec
tions
Contact specialist dealer
Seat bottom LED is flashing Seat tilt motor fault Faulty cabling or plug con
tact
Defective actuator
Check cabling/plug connec
tions
Contact specialist dealer
2 LEDs below the seat bot
tom are flashing
Seat height adjustment motor
fault
Faulty cabling or plug con
tact
Defective actuator
Check cabling/plug connec
tions
Contact specialist dealer