5.4 Installation Instructions
This equipment (sensor unit) is connected to the
piping with high pressure cone & thread tting, 9/16"
dia. 562C. (Connection examples are shown in Figs
5.3 and 5.4.)
Connector: High pressure cone & thread connection,
9/16" dia. 562C (male thread 11/8 - 12UNF)
Set the collar so that the groove faces out.
5.5 Sensor Unit-to-Piping Connection Guidelines
5.5.1 Machining the Female Connector (for reference)
Fig. 5.5 is a machining drawing for preparing the high pressure cone & thread connection, female
connector. See also Fig. 5.6 "Detail Drawing of Connectors".
Fig. 5.3
Sensor Unit (body)
Sensor Unit (connector)
Piping (Connectors)
Piping
Fig. 5.4
M6 Hex Soc. Hd. Bolts
Connector
Clamp
9/1618UNF LH
Gland Nut
Collar
Fig. 5.5
To the Meter
(Sensor Unit)
Weep Hole φ2.34
(2 places)
φ6.5 max. must
be kept strictly.