The following settings are available:
Designation Description
Storage rate Select storage interval of the measurement
(0.1 ms*, 1 ms, 5 ms, 10 ms, 100 ms, 500ms,
1s, 5s, 10s, 30s, 1min, 5min, 10min, 30min)
Storage time Select storage time for the measurement
(10s, 30s, 1min, 5min, 10min, 30min, 1h,
12h, 24h, 3d, 7d).
When the storage time has expired, the measure-
ment stops automatically
Pre-Trigger time Select the period in which the measured values are
already recorded before the trigger is activated
(OFF, 5s, 10s, 30s, 1min, 10min, 30min)
Trigger Activate the Trigger function and define settings for
the trigger signal.
► Further information is available in Chapter & „Set
trigger“ on Page54.
Ring buffer If this option is activated, the defined memory area is
continuously overwritten by current measured data
Auto repeat When the option is activated, the measurement is
started again the next time the trigger is triggered af-
ter automatic storage of the previous measurement
File name Name of the measurement recording
*only applies to AUX 1; all other channels continue to be sampled at 1 ms.
53
The Parker Service Master COMPACT V1.0/02/24
Operation