Maintenance
95
Figure 20 White ceramic connector
5. Remove the lamp assembly from the bracket by slackening the thumbscrew
and pulling the lamp mount vertically upward.
Save the thumbscrew for use with the new lamp assembly.
6. Unpack the new lamp assembly, taking care to hold it only by the metal mount
to prevent fingermarks on the bulb.
7. Slip the slot at the base of the lamp mount over the stud on the bracket in the
lamp compartment and then secure with the thumbscrew.
8. Carefully push the ceramic connector firmly onto the pins on the base of the
lamp.
9. Wipe the bulb with a soft cloth moistened with alcohol to remove dirt, since
this would otherwise be burned in when the lamp is hot.
10. Replace the lamp baffle using the reverse of the procedure described in step 3.
11. Replace the lamp compartment cover.
This completes the halogen lamp replacement procedure.
Thumbscrew
Halogen lamp
assembly
White ceramic
connector
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 25 and is the answer not in the manual?

PerkinElmer Lambda 25 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 25
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals