PATIENT MONITORING
SERVICE GUIDE
Measurements and Monitoring
M3/M4 Monitors
(M3046A)
Measurement Server
(M3001A and M3000A)
Measurement Server Extensions
(M3015A and M3016A)

Table of Contents

Questions and Answers

  • M
    Mrs. Kimberly GentrySep 12, 2025
    What to do if Philips Measuring Instruments show 'BAD SERVER LINK'?
    • J
      james88Sep 12, 2025
      If your Philips Measuring Instruments display 'BAD SERVER LINK', it means that you cannot use this combination of Monitor, Measurement Server and cable. Switch off the monitor and contact your biomedical department.
  • D
    Dominic TaylorSep 14, 2025
    How to resolve 'BAD SERVER LINK plus Measurement Server Revision not supported' on Philips M3046A?
    • K
      Kelsey SmithSep 14, 2025
      If your Philips Measuring Instruments display 'BAD SERVER LINK' and the status message 'Measurement Server Revision not supported' in red, it means an M3000A Measurement Server with revision A software is connected to an M3046A Monitor with a software revision of B or later, a combination that does not allow monitoring. Update the software to resolve this issue.

Summary

1 Introduction to the Instrument

Theories of Operation and Functional Descriptions

Details on monitor, server, and extension theories and functions.

2 Installing the Instrument

Safety

Patient safety, leakage current, and protection against electric shock.

3 Maintaining the Instrument

Recommendations for Maintenance Frequency

Timetable for suggested maintenance activities for various components.

Preventive Maintenance Tasks

Specific maintenance tasks for Monitor, Measurement Server, and Extensions.

4 Testing the Instrument

Accuracy and Performance Procedures

Detailed procedures for testing temperature, ECG/Resp, invasive pressure, etc.

Patient Safety Checks

Procedures for verifying system enclosure leakage current and protective earth continuity.

5 Troubleshooting the Instrument

Part 2 Isolating and Solving Instrument Problems

Detailed troubleshooting for INOP messages, defective components, and specific problems.

6 Repairing the Instrument

Disassembly Procedures for the M3015 A Measurement Server Extension

Steps for removing covers, pump, and CO2 scrubber from the extension.

Refit Procedures for the M3015 A Measurement Server Extension

Steps for refitting the CO2 scrubber and pump into the extension.