PAGE 9-51097626, VER. 03
18. Once the device has completed the code load and rebooted proceed to step 19.
19. The data recovery tool should now detect the device and begin downloading the data
a. Verify the correct LPC software version was loaded. The Usage Hours Tab will indicate which
software version is to be installed.
20. Put the device in Demo mode by clicking the Demo mode check box on the Real Time Data tab
21. Verify that the word DEMO appears on the LCD screen
22. Select Pulse Mode and a setting of 2.0
23. Verify that the Battery symbol appears in the display window. Record results
24. Verify that the External Power Status symbol appears in the display window. Record results.
25. Observe the Pulsemeter and verify that the SimplyGo is automatically triggering breaths. Record
results.
26. Verify that the Pulse Symbol appears in the display window when a pulse is triggered. Record
results.
27. Block the Pulsemeter exhaust port, wait for the SimplyGo to trigger a breath and observe the read-
ing on the Pulsemeter scale. Immediately unblock the exhaust port to exhaust the Pulsemeter and
block it again before the next pulse. Perform this for the next 6 consecutive pulses and record the
results on the SimplyGo Pulse Flow Measurements Worksheet to calculate the Average. Record
the Average reading on the Final Test Data Sheet.
28. Select Pulse Mode and a setting of 6.0. Repeat step 27.
29. Select Sleep Mode
30. Perform step 27 for a setting of 2.0.
31. Perform step 27 for a setting of 6.0.
32. Disconnect the Pulsemeter from the patient cannula connector.
33. Select Continuous Mode and a setting of 0.5.
34. Remove the Pulsemeter tubing from the pulsemeter and connect one end of the tubing to the bot-
tom of the flowmeter.
IMPORTANT NOTE
Do NOT turn the device off/on during the programming.
When programming is complete, the device will reboot.