PAGE 9-6
1097626, V
ER. 03
35. Connect the flowmeter to the patient cannula connector.
36. Holding the Flowmeter in the vertical position above the SimplyGo, observe the Flowmeter scale
and record the reading on the Test results sheet.
37. Repeat step 36 for a setting of 2.0 continuous flow.
38. Disconnect the Flowmeter from the patient cannula connector.
39. Connect a calibrated O2 Analyzer to the Patient Cannula connector as shown below
40. Perform steps a and b.
a. Let the device run for at least 10 minutes on 2 LPM continuous flow. Record the results of the
O2 purity from the O2 Analyzer and the internal O2 sensor (Real Time Data Screen) on the
test data sheet.
IMPORTANT NOTE
If performing both the Final Test and Balancing Procedure, DO NOT perform step 40 a and b. Continue
directly to the Balancing Procedure.