PAGE 5-211097626, VER. 03
High Pressure failure
• Check for kinked tubing
• Check Patient Filter for
obstructions
• Main PCA
• Use Real time data and confirm Event
Code
• Remove Obstructions, replace Patient
Filter
• Replace Main PCA
Excessive Number of Motor
Restarts Failure
• Compressor
• Wire connections to the
Hall effect sensor
• Check the connections
• Replace the compressor
Power Off Due To No
Breathing for 30 minutes in
Day Time mode
• Power off • Normal event
Power Off Due To a Technical
Failure
• Main PCA
• Valves
• Fan
• Replace the Main PCA
• Check wire connections
PROBLEM / EVENT CODE POSSIBLE CAUSE RECOMMENDED ACTIONS