PAGE 7-231097626, VER. 03
7.11 COMPRESSOR COVER REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. Remove the Front Enclosure.
3. Remove the Sieve Canisters to gain access to the Compressor Cover screws.
4. Remove the six philips screws that secure the compressor cover to the spine assembly.
Installation: (for item identification, refer to removal instructions above)
1. Place the Compressor Cover over the Compressor in the spine assembly.
2. Secure the Compressor Cover to the spine assembly by securing the six screws.
3. Reinstall the Sieve Canisters.
4. Install the Front Enclosure.
5. Install the Rear Enclosure.