PAGE 7-22
1097626, V
ER. 03
Installation: (for item identification, refer to removal instructions above)
1. Place the two o-rings into the spine assembly.
2. Install the Sieve Bed Balance Valve with wire harness towards the opening in the spine assembly.
3. Secure the Sieve Bed Balance Valve to the spine assembly by tightening the two screws.
4. Thread the wire harness through the hole in the spine assembly and connect it to J14 on the Main
PCA.
5. Install the Front Enclosure.
6. Install the Rear Enclosure.