PAGE 7-311097626, VER. 03
Installation: (for item identification, refer to removal instructions above)
1. Place the compressor into the spine assembly.
2. Connect the Compressor power and Hall Effect sensor connections to the Main PCA.
3. Connect the intake hose to the Compressor intake manifold.
4. Connect the Compressor outlet hose to the valve hose elbow.
5. Secure the Compressor mount bracket to the spine assembly by tightening the four screws.
Repeat this process for the second Compressor mount bracket.
6. Install the Compressor Cover.
7. Install the Sieve Canisters.
8. With the SimplyGo sitting in the upright position, place the Alignment jig on top of the spine.
9. Loosen the outlet tube using two wrenches and then allow the metal tube to sit in the jig.
10. Tighten the clamp to finish the connection at the other end of the metal outlet tube.
11. Slide the Alignment jig off the unit and set it aside.
NOTE
It is recommended to use the Compressor Alignment Tool (Respironics Part Number: 1102513) any time a
technician is installing a compressor.
NOTE
The springs will have to be stretched out to align the holes in the mounting bracket wit the holes on the
spine assembly.