12. Before the unit is completely reassembled use the Compressor tube gauge to ensure proper spac-
ing between the tubes. The tubes should only just touch the gauge when the tubes are aligned
correctly.
13. Install the Front Enclosure.
14. Install the Rear Enclosure.
SimplyGo Compressor Upgrade Programming Procedure
1. Download and install the SimplyGo/SimplyFlo Data Recovery Tool and Manifest File. Follow the
procedure outlined in Chapter 5.3 sections 5.3.1 through 5.3.3.
2. Open the Data Recovery Tool application by double clicking on the icon below.