Please note that the calibration solutions are not provided by the manufacturer.
Read the signal value on the display of the photobioreactor control unit (Fig. 17B), wait until it is stable and confirm
this value by pressing the “S” key.
Repeat the same procedure with buffer solution of pH = 10.
When calibration is done, press the “M” key to get back to the main menu.
Optional: Calibration accuracy may be verified by plunging the electrode in the same calibration solutions of
known pH (pH value in between 7 and 10).
Fig. 17 pH sensor calibration. A) Electrode is placed into the solution of pH 7. B) Reading of the measured value. C, D, E) Grounding with
the copper cable for the calibration of InPro3253/120/PT1000 version of the electrode.