Description
1
Table
1-1
LIFEPAK
9
defibrillator/monitor
Specifications*
(cont.)
General
AC
INPUT
OPTIONS
120
or
240Vac
nominal
line
voltage
without
adjustment.
50
or
60Hz
with
adjustment
in
Setup
Menu.
BATTERY
TYPE
Sealed
lead-acid,
16Vdc
nominal,
3
Ahr.
BATTERY
CAPACITY
A
new,
fully-charged
battery
will
provide
one
of
the
following
prior
to
shutdown:
Typical
Minimum
Number
of
360J
discharges:
75
40
Minutes
of
monitoring:
90
70
LOW
BATTERY
INDICATOR
Advises
operator
to
connect
ac
power.
Battery
capacities
before
and
BATTERY
CHARGE
INDICATOR
BATTERY
CHARGE
TIME
SERVICE
INDICATOR
POWER
CONSUMPTION
SIZE
Height:
Width:
Depth:
Weight:
POWER
CORD
LENGTH
STANDARD
PADDLE
ELECTRODE
AREA
after
warning
(LOW
BATTERY-CONNECT
AC
POWER)
for
a
new,
fully
charged
battery
are:
Typical
Minimum
Number
of
360J
discharges
before
warning:
50
20
Number
of
360J
discharges
after
warning:
25
5
Indicator
BATT
CHRG
illuminates
when
battery
is
charging.
24hrs
to
full
capacity.
Message
SERVICE
indicates
self-diagnostic
routines
have
detected
improper
operation
requiring
service
attention.
160W
maximum
while
monitoring
with
recorder
on
and
defibrillator
charging.
35.2cm
(13.9in)
29.7cm
(11.7in)
31.0cm
(12.2in)
12.7kg
(28lbs)
3m
(10ft)
82cm?
Environmental
ATMOSPHERIC
PRESSURE
RELATIVE
HUMIDITY
TEMPERATURE
RANGE
797
to
500mm
Hg
(-570
to
11,000ft).
0 to
95%
(non-condensing)
at
0
to
34°C
(32
to
94°F).
0 to
80%
(non-condensing)
at
35
to
45°C
(95
to
113°F).
0
to
45°C
(32
to
113°F)
operating.
—30
to
65°C
(-22
to
149°F)
storage.
©
August
1993
Physio—Control
All
Rights
Reserved
1-5