LIFEPAK
9
defibrillator/monitor
Warning
POSSIBLE
ARCING
AND
PADDLE
DAMAGE.
When
discharging
the
defibrillator
into
the
internal
test
load,
make
sure
the
standard
paddles
are
securely
and
properly
stored
in
the
paddle
storage
area
(STERNUM
paddle
on
the
left,
APEX
paddle
on
the
right).
This
helps
prevent
arcing
and
formation
of
pits
on
paddle
electrode
surfaces.
Pitted
or
damaged
paddle
electrode
surfaces
can
cause
patient
skin
burns
during
defibrillation.
PIP-General
Instructions
The
following
sections
are
the
actual
PIP
steps
to
be
performed.
These
steps
are
designed
to
be
executed
in
sequential
order.
For
instance,
if
a
heart
rate
input
of
30bpm
is
specified
for
one
test
it
also
applies
to
the
next
test
unless
you
are
instructed
to
change
it.
PIP-Physical
Inspection
Before
beginning
the
PIP-Physical
Inspection,
be
sure
to
disconnect
the
ac
power
cord
from
the
ac
power
source.
e
Inspect
all
exterior
surfaces
for
signs
of
abuse
or
other
damage,
including
fluid
spills,
cracks
and
dents.
e
Inspect
paddles
for
pitted
electrode
plates,
presence
of
dried
defibrillation
gel,
and
integrity
of
cables
and
switches.
Remove
gel
or
other
foreign
matter
from
paddles
and
paddle
wells.
e
Inspect
recorder
for
ease
of
door
operation
and
positive
latching.
Remove
dust
or
foreign
matter
inside
recorder.
e
Inspect
all
external
connections
for
bent,
broken,
or
corroded
pins,
and
debris.
e
Inspect
the
power
cord
and
all
cables
and
accessories
for
damage,
cracks,
breaks,
bent
or
broken
connector
pins,
or
other
signs
of
wear.
e
Open
the
battery
door
on
the
bottom
of
the
defibrillator/monitor
and
make
sure
batteries
are
installed.
Close
the
door
and
make
sure
it
latches
properly.
e
©
Press
the
ON
pushbutton
and
confirm
that
the
green
LED
illuminates.
Press
ON
again
and
make
sure
the
green
LED
is
off
and
power
is
off
(the
display
is
blank).
Then
press
all
other
pushbuttons
and
check
them
for
uniform
mechanical
resistance.
PIP-AC
Power
1.
Connect
the
defibrillator/monitor
to
ac
power.
2.
Ifthe
dash
number
of the
defibrillator/monitor
is
listed
below,
confirm
that
the
power module
clip
is
installed
at
the
bottom
of
the
rear
panel
power
switch,
locking
the
switch
to
the
power
on
(I)
position.
Then
go
on
to
step
3.
Dash
Numbers
with
Power
Module
Clip
Installed
803800-09
803800-300
803800-
18
803800-309
thru
-312
803800-28
803800-314
thru
-316
170008-000
A
3-4
All
Rights
Reserved
O
August
1993
Physio—Quntrol

Table of Contents

Other manuals for Physio Control LIFEPAK 9

Related product manuals