PicoQuant GmbH HydraHarp 400 Software V. 3.0.0.1
TTTR Marker Settings Group
Here you can set the signal specifications for external marker signals used in TTTR mode (T2 and T3 mode).
The settings for the different markers are independent from each other. The following table applies to each of
the four markers:
enable Check this if you want the marker to be included into the TTTR data
stream.
rising / falling Radio buttons to identify the active edge of the signal.
Reference Clock Source Group
The setting in this group takes effect after the next software start.
internal / external Select whether the HydraHarp uses its internal clock or an external 10 MHz
clock supplied at the REF IN connector.
6.8. About HydraHarp Dialog
This dialog provides version information on the HydraHarp software and hardware, the latter only if it is
installed an operational. It can be opened via the Help menu or via the toolbar. The button Request Support
opens a small text viewer window that provides a listing of HydraHarp hardware and software versions that you
can copy and paste into your support enquiry. Note that this information is very important for adequate support.
Support requests without this information cannot be processed and will be delayed by return questions for this
information. If your system is not functional at all, the minimum information you must provide for support is the
serial number of your HydraHarp. It can be found at the back of the housing. The dialog also provides buttons
for links to relevant web pages and a button for download of software updates. Note that downloading a
software update does not automatically install it. Typically the downloads contain zip files that must be
unpacked and installed manually. Also note that it may not always be advisable to blindly install the latest
software. This is the case especially when a user or his/her laboratry have developed custom software that
relies on a certain file format. Updates may break such compatibility. Therefore, always study the release notes
before installing.
6.9. Title and Comment Editor
You can use this dialog to edit the file title / comment. It can be opened via double–click on the title, via the File
menu or via the Print Preview Toolbar. The text you enter here will be stored in the data file. The first line will
be displayed as the file title over the histogram display area. The text you can enter here is limited to 4 lines
and 255 characters. Upon loading of a data file, the title and comment will also be retrieved. It will also be
included in prints, if the corresponding check mark is set in the page setup dialog box available through the
File menu.
6.10. Print Preview Dialog
Use this Dialog to preview the layout as it would appear when printed. When you choose this command, the
main window will be replaced with a print preview window in which one or two pages will be displayed in their
printed format. The print preview dialog offers the following options:
Print
Bring up the print dialog box, to start a print job.
Next Page
Preview the next printed page.
Prev Page
Preview the previous printed page.
One Page / Two Page
Preview one or two printed pages at a time.