PRS 500 E
Instructions for use
5069-0-8002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 von 47
CAUTION!
The instructions for use contain all the information relevant to safety in order
to generally put the X-ray system into operation. The device may only be
operated by appropriately trained and trained personnel. In this context,
operation is ensured by clear symbols on the control elements. All further
information and instructions can be found on the supplied data carrier (USB,
CD or DVD). This information applies in its entirety as an appendix to these
instructions for use and must be observed.
NOTE
All control elements are marked with clear symbols on the control console and on
the swivel arm or the image receptor stand, which are described again in detail in
the corresponding instructions for use. The legal requirements regarding the
building regulations for an X-ray area must be met. The X-ray system must be tested
in accordance with the regulations in force in the country of installation and
approved by the appropriate body.
CAUTION!
If the wrong SID is set for an exposure, it can have a damaging effect on the
patient. The inverse square law applies. Halving the distance leads to a
radiation dose that is four times higher.
WARNING!
X-rays may not be performed on persons without a medical justifying
indication. In the case of pregnant women and children, it must be carefully
considered whether an exposure is necessary. It should be avoided if possible.